Technical Support Information No. 272

Technical Support Information No. 272
Technical Support Information No. 272
                                                                                     Revision 1.2         Aug 2018
Process Industries and Drives Division – Process Automation – Analytical Products and Solutions                                           page 1/10




SITRANS SL - Exchange of the ATEX cables
Only qualified service technicians who are experienced with the requirements for Ex d and Ex e
and who are able to evaluate the result are allowed to exchange the ATEX cables!




Follow also the safety notes in the Operating Instruction A5E01132948-04
and in the Certificate PTB 08 ATEX 1008X

Required Tools:
Allen key 2.5mm, small screw driver, flat spanners with wrench sizes 22, 24, 27, 30 and 36mm

Dismounting of the existing cables

     ·    Switch off the SITRANS SL in the Ex zone and wait for 2 minutes
     ·    Switch off the sensor purging and dismount it from the sensor
     ·    Dismount the sensor connection cable between the junction boxes and transport the sensor heads to a clean
          area
     ·    There may be some white powder in the junction boxes: this is talcum powder and should be removed
     ·    The window flanges stay on the process and also the process purging stays switched on: the process keeps
          closed and the windows keep clean, also the alignment is not affected

In clean area:

     ·    Open the safety screws on the lids and open the sensor heads
     ·    Dismount carefully the transmitter and receiver loop cables and main connection cable from the NPT thread
          in the sensor housings and the junction boxes. In case of connection cable analog, the reducing adapter
          must not be removed.

     Note: Check that the NPT threads of the sensor housings are complete and undamaged

Unrestricted © SIEMENS AG 2018. All rights reserved.                                                                                 A5E44986296A/003

While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full
agreement. Technical data are subject to change.
Technical Support Information No. 272
Technical Support Information No. 272
                                                                                     Revision 1.2         Aug 2018
Process Industries and Drives Division – Process Automation – Analytical Products and Solutions                                           page 2/10




     Mounting Transmitter Sensor
  · Screw the ATEX cable gland of the transmitter junction box cable with 2 windings teflon tape into the transmitter
    housing as shown below.

  ATEX Cable gland                                                                Wrap the wire ends with isolating tape to a strand
                                                                                 to make it easier to insert the cable and to avoid to
                                                                                         damage the isolation of the wires.

                                                                                   Release the two nuts (1) and (2) a bit to screw in
                                                                                    the cable gland without turning the complete
                                                                                                       cable.

                                                                                   After mounting the cable gland tighten the nuts
                                                                                           (before connecting the wires)




                                                                                                                      2           1




             Transmitter                                              Teflon tape
              housing

            According ATEX and FM standard at least 5 threads have to be screwed in.


  · Connect the wires according the following table inside the transmitter housing:



                                                                                                                              Connection
                                                                                         Colour Code
                                                                                                                               position
                                                                                              Red                                S1
                                                                                              Blue                               S2
                                                                                              Pink                               S3
                                                                                              Grey                               S4
                                                                                             White                               S5
                                                                                             Brown                               S6
                                                                                             Green                             Chassis
                                                                                             Yellow                            Chassis
                                                                                             Shield                            Chassis



Unrestricted © SIEMENS AG 2018. All rights reserved.                                                                                  A5E44986296A/003

While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full
agreement. Technical data are subject to change.
Technical Support Information No. 272
Technical Support Information No. 272
                                                                                     Revision 1.2         Aug 2018
Process Industries and Drives Division – Process Automation – Analytical Products and Solutions                                           page 3/10



  · Mount the Transmitter junction box cable at the Transmitter junction box as shown below. Connect the
    shield on the cable gland.
                                                                                      Transmitter junction
                                                                                           box cable



                                     Junction box




  · Connect the wires according the following table

                                                Wire colour                          Connection position
                                                   Red                                        1
                                                   Blue                                       2
                                                   Pink                                       3
                                                   Grey                                       4
                                                   White                                      5
                                                  Brown                                       6
                                                  Green                                  PE terminal
                                                  Yellow                                 PE terminal
                                                  Shield                                   Gland




Unrestricted © SIEMENS AG 2018. All rights reserved.                                                                                 A5E44986296A/003

While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full
agreement. Technical data are subject to change.
Technical Support Information No. 272
Technical Support Information No. 272
                                                                                     Revision 1.2         Aug 2018
Process Industries and Drives Division – Process Automation – Analytical Products and Solutions                                           page 4/10




Mounting Receiver Sensor

  · Screw the ATEX cable gland of the receiver junction box cable with 2 windings Teflon tape into the receiver
    housing as shown below.


                                                                                    Wrap the wire ends with isolating tape to a strand
                                                                                   to make it easier to insert the cable and to avoid to
ATEX Cable gland
                                                                                           damage the isolation of the wires.

                                                                                     Release the two nuts (1) and (2) a bit to screw in
                                                                                      the cable gland without turning the complete
                                                                                                         cable.

                                                                                     After mounting the cable gland tighten the nuts
                                                                                             (before connecting the wires)



                                                                                                                                   2           1


                                                                   Teflon tape

                  Receiver
                  housing
            According ATEX and FM standard at least 5 threads have to be screwed in.




  · Connect the wires of the Receiver junction box cable inside the receiver housing according the following table

                                                                                          Wire colour                         Connection
                                                                                                                               position
                                                                                              Red                                L1
                                                                                              Blue                               L2
                                                                                              Pink                               L3
                                                                                              Grey                               L4
                                                                                             White                               L5
                                                                                             Brown                               L6
                                                                                             Green                               16
                                                                                             Yellow                              17
                                                                                             Black                               18
                                                                                             Violet                              19
                                                                                             Shield                            Chassis




Unrestricted © SIEMENS AG 2018. All rights reserved.                                                                                 A5E44986296A/003

While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full
agreement. Technical data are subject to change.
Technical Support Information No. 272
Technical Support Information No. 272
                                                                                     Revision 1.2         Aug 2018
Process Industries and Drives Division – Process Automation – Analytical Products and Solutions                                           page 5/10




  · Connect the Receiver junction box cable at the Receiver junction box as shown below. Connect the shield
    on the cable gland.
                                         Receiver junction
                                            box cable


                                        Junction box




  · Connect the wires according the following table


                                                Wire colour                          Connection position
                                                   Red                                       1
                                                   Blue                                      2
                                                   Pink                                      3
                                                   Grey                                      4
                                                   White                                     5
                                                  Brown                                      6
                                                  Green                                      8
                                                  Yellow                                     9
                                                   Black                                     10
                                                   Violet                                    11
                                                  Shield                                   Gland




Unrestricted © SIEMENS AG 2018. All rights reserved.                                                                                 A5E44986296A/003

While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full
agreement. Technical data are subject to change.
Technical Support Information No. 272
Technical Support Information No. 272
                                                                                     Revision 1.2         Aug 2018
Process Industries and Drives Division – Process Automation – Analytical Products and Solutions                                           page 6/10




Mounting and Connection of the Receiver Connection Cable (Analog /
MODBUS)
  · If the reducing adapter Ex d 1/2NPT / 3/4NPT must be reassembled, screw it with 2 windings Teflon tape into
    the receiver housing as shown below.




                                                                                                                         Reducing adapter Ex d
                                                                                                                            1/2NPT / 3/4NPT




                                                                                                                     Teflon tape




            According ATEX and FM standard at least 5 threads have to be screwed in.

  · Screw the ATEX cable gland of the Analog or Modbus Connection cable with 2 windings Teflon tape into
    the reducing adapter as shown below

                                     Wrap the wire ends with isolating tape to a strand to make it easier to insert the
                                                cable and to avoid to damage the isolation of the wires.

                                    Release the two nuts (1) and (2) a bit to screw in the cable gland without turning
                                                                  the complete cable.

                                        Mark the cable gland to count the threads. Screw the cable glands into the
                                       reducing adapter with 5-6 threads: hand-tight plus 1/6 thread (1 spanner flat)

                                                           After mounting the cable gland tighten the nuts.

                                  Reducing adapter




                                                                 2            1




Unrestricted © SIEMENS AG 2018. All rights reserved.                                                                                 A5E44986296A/003

While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full
agreement. Technical data are subject to change.
Technical Support Information No. 272
Technical Support Information No. 272
                                                                                     Revision 1.2         Aug 2018
Process Industries and Drives Division – Process Automation – Analytical Products and Solutions                                           page 7/10




                                                                          After mounting the cable gland tighten the nuts.




  · Before connecting the wires: check with a multimeter that there is no contact between the cable bundle and
    ground: measure between the cable gland and the single wires that there is no connection.
      Check all wires!




Unrestricted © SIEMENS AG 2018. All rights reserved.                                                                                 A5E44986296A/003

While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full
agreement. Technical data are subject to change.
Technical Support Information No. 272
Technical Support Information No. 272
                                                                                     Revision 1.2         Aug 2018
Process Industries and Drives Division – Process Automation – Analytical Products and Solutions                                           page 8/10




  · Connect the wires of the Receiver connection cable inside the Receiver housing according the following table.
    Note: there is a not used terminal between position 2 and 3!


                                                                                                                              Connection
                                                                                          Wire colour
                                                                                                                               position
                                                                                              Red                                 1
                                                                                              Blue                                2
                                                                                              Pink                                3
                                                                                              Grey                                4
                                                                                             White                                5
                                                                                             Brown                                6
                                                                                             Green                                7
                                                                                             Yellow                               8
                                                                                             Black                                9
                                                                                             Violet                              10
                                                                                           Grey-Pink                             11
                                                                                           Red-Blue                              12
                                                                                         White-Green                             13
                                                                                         Brown-Green                             14
                                                                                         White-Yellow                            15
                                                                                          White-Grey                             20
                    Receiver of SITRANS SL 1.1                                            Grey-Brown                             21
                                                                                          White-Pink                             22
                                                                                          Pink-Brown                             23
                                                                                         Yellow-Brown                          Chassis
                                                                                             Shield                            Chassis




Unrestricted © SIEMENS AG 2018. All rights reserved.                                                                                 A5E44986296A/003

While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full
agreement. Technical data are subject to change.
Technical Support Information No. 272
Technical Support Information No. 272
                                                                                     Revision 1.2         Aug 2018
Process Industries and Drives Division – Process Automation – Analytical Products and Solutions                                           page 9/10




Mounting and Connection of the Receiver Connection Cable (PROFIBUS)

          According ATEX and FM standard at least 5 threads have to be screwed in.

  · Screw the ATEX cable gland of the Profibus Connection cable with 2 windings Teflon tape into the receiver
    housing as shown below


                        Wrap the wire ends with isolating tape to a strand to make it easier to insert the
                                   cable and to avoid to damage the isolation of the wires.

                       Release the two nuts (1) and (2) a bit to screw in the cable gland without turning
                                                     the complete cable.

                                   Screw the cable gland into the receiver housing with 5 threads.

                                              After mounting the cable gland tighten the nuts.




                                    2
                                                     1




              Teflon tape
                                                                                    Receiver housing

                                                                After mounting the cable gland tighten the nuts.

  · Before connecting the wires: check with a multimeter that there is no contact between the cable bundle and
    ground: measure between the cable gland and the single wires that there is no connection.
    Check all wires! There is only transmission between the shield wires and ground.




Unrestricted © SIEMENS AG 2018. All rights reserved.                                                                                 A5E44986296A/003

While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full
agreement. Technical data are subject to change.
Technical Support Information No. 272
Technical Support Information No. 272
                                                                                     Revision 1.2         Aug 2018
Process Industries and Drives Division – Process Automation – Analytical Products and Solutions                                           page 10/10




  · Connect the wires of the Receiver connection cable inside the Receiver housing according the following table.


                                                                                                                              Connection
                                                                                         Wire Colour
                                                                                                                               position
                                                                                           Black (1)                              1
                                                                                           Black (2)                           Chassis
                                                                                           Black (N)                             13
                                                                                           Black (P)                             14
                                                                                           Black (3)                              2
                                                                                           Black (4)                           Chassis
                                                                                         Black (Shield)                          15




Finalizing the assembly

     ·    Mount and tighten the lids
     ·    Take care that no wire is squeezed in the thread and the isolation is not damaged
     ·    Lock the lids with the safety screw against opening
     ·    Mount the sensor heads on the window flanges
     ·    Connect the sensor connection cable to the junction boxes
     ·    Connect the sensor purging and switch it on
     ·    Switch on the analyzer in Ex zone only when all lids and junction boxes are closed
     ·    Check operation of the analyzer and the customer interface

Note: All SITRANS SL components have to be connected to the equipotential bonding.




Unrestricted © SIEMENS AG 2018. All rights reserved.                                                                                 A5E44986296A/003

While we have verified the contents of this TSI in agreement with hardware and software described, variations remain possible. Thus we cannot guarantee full
agreement. Technical data are subject to change.