2016 NEQAS-CC National External Quality Assessment Scheme in Clinical Chemistry - LUNG CENTER OF THE PHILIPPINES
←
→
Page content transcription
If your browser does not render page correctly, please read the page content below
LUNG CENTER OF THE PHILIPPINES
National Reference Laboratory for Clinical Chemistry
2016 NEQAS-CC
National
External
Quality
Assessment
Scheme in
Clinical ChemistryDisclaimer
The presentation slides used for this Orientation
Workshop on 2016 NEQAS-CC Monthly Program, and its
related LCP-NRL- CC documents containing the “RANDOX-
RIQAS 9128 monthly program” are intended ONLY for the
purpose of providing the 2016 National External Quality
Assessment Scheme and are NOT to be considered or taken
as a form of endorsement or advertisement.
The clinical laboratories are cautioned that this
presentation DOES NOT constitute an offer, advice, invitation,
or recommendation to purchase or subscribe to any of the
RANDOX products. To give a brief description of the
NEQAS-CC Program for 2016
To update the clinical laboratory
participants about the changes in
the 2016 NEQASS-CC Program
2016 NEQAS-CC
Objectives To assist clinical laboratory
participants on how RIQAS online
is used with regards to:
Registration
Configuration
Submission of Results On-line
Troubleshooting2016 NEQAS-CC Monthly Program
PROCESS FLOW
NEQAS Registration of Contract and
Bidding Participants NEQAS sample
Process (April-May: DM orders
(Jan-Mar ) deadline) (June)
SUBMISSION
RANDOX)
of Monthly Start of the ORIENTATION
Results NEQAS-CC OF
via on-line Program PARTICIPANTS
(*July-June) (August: Testing of (July-August)
www.riqas.net) the samples)
STATISTICAL DISCLOSURE SELF-
ANALYSIS of Monthly and End EVALUATION
(RIQAS 9128) of Cycle NEQAS- -Review of reports
By Randox CC Reports to -Perform corrective
Participants action if neededChanges:
Cycle
Delivery of Samples
2016 NEQAS-CC
Monthly Program
Schedule of Deadlines
Randox- Reports and
RIQAS9128
Guidelines
Certification One year cycle =
12 monthly Cycle
RIQAS 9128 consists of:
Cycle Cycle 13B –
July to December 2016
Cycle 14 A –
January to June 20172 batches:
Delivery 1st batch of delivery
Cycle 13B
of July-August 2016
Samples 2nd batch of delivery
Cycle 14A
November 2016Stability-
Lyophilized
Once reconstituted,
viable for 2 days
2016
NEQAS-CC
Samples Deadline is last Monday
of the Month
“Late result” will NOT be
Submission accepted and included in the
of Results statistical data, an email will
be sent giving the average of
Schedule the comparator group instead
of
Deadlines Please DO NOT beat the
deadline mentioned so as to
have an allowance for any
changes Corrected results may
be accepted, up to 4
weeks after the final date,
Submission but ONLY for
transcription errors
of Results
A lab wishing to submit a
corrected result must
Schedule
complete a request form
of
Deadlines and submit the analyser
print out as an evidence
that a transcription error
was made All results are processed by
RIQAS within 24 – 72 hours
of the Final Date deadline
Reports Based on Standard
and Deviation
Guidelines Participants can quickly
appraise their own
performance, in comparison
with that of other laboratory
participantsReports and Guidelines
Text Section: Levey Jennings Chart:
Contains all the Shows performance
statistical with time
information Target Scoring Chart:
Provides a
visualization of a
Lab’s Target Scores
Histogram: % Deviation by
Provides a method Sample Chart:
and instrument Helps to identify
trends and shifts in
comparison
performance
Multi Method Stat % Deviation by
Section: Concentration Chart:
Enables assessment Provides a rapid
of the performance assessment of
of each method concentration
related biasesSolid vertical line:
Change in registration
Method
Instrument
Reagent supplier
Unit of
measurementText Section N: Number of results used to generate the mean Mean: The average result, based on a consensus CV%: Coefficient of Variation Um: Uncertainty of the Mean SDPA: Standard Deviation for Performance Assessment Exc.: Excluded results (outliers)
This contains a summary of
your performance
―Your Result
― Mean for Comparison
― Indicators of
Performance
Standard Deviation
Index (SDI)
Target Score (TS)
% Deviation (%DEV)
Various Running
Mean Indices (RM)Histogram:
Provides group comparison
All Method Group
Your Method Group
Your Instrument Group
Your ResultTarget Score Chart
Excellent
Good
Acceptable
Need for
improvement
Unacceptable
Target Score
Sample
Number This provides a numerical index of
performance and allows participants
to assess performance at a glance
Target
Target Score (TS) Interpretation:
101-120 Excellent
Score 71-100 Good
51-70 Acceptable
41-50 Needs Improvement
10-40 Unacceptablea black dot within the white section
represents results within the acceptable
limits of performance
The black dots represent the % Dev of each reported
result
The white circles represent the
RM%Dev at each time point, and
will help to identify any distinct
trends
And a black dot within one of the red sections
represents results outside the acceptable
limits of performance% Deviation by Concentration Chart:
Current sample
indicated by the square
The circles represent previous
samples at different concentrations check first the Summary page
(located at the back of the RIQAS
report) to assess your performance
for the most recent sample
Look for red triangle ( ),
focus your investigations on these
analytes
Remember
The triangle will only appear when
all limits are exceeded:
SDI >2
Target Score < 50
% Deviation > Acceptable
limits of performanceFor the LJ and Target score charts:
Shading indicates performance,
The lighter the shading, the
more accurate the results
Remember
% Deviation, which indicates by
what percentage the
laboratory’s result deviates from
the mean for comparisonRequirements:
Certificate of at least
Participation four (4) sample results
to the for cycle 13B, and
2016 NEQAS four (4) sample results
for cycle 14ACertificate of
Participation At least six (6) out of
on Randox- the 12 samples
Riqas AO 2007-0027:
Secondary Category : Routine
Clinical Chem – Gluc, BUN, UA,
Crea, Total Cholesterol
Please do not beat the
Remember
deadline given so as to have
allowance for any changes
Always update your e-mail
account and the contact
personSummary
Cycle ONE year, 12 mos (Cycle 13B and Cycle 14A)
Delivery of Samples 2 batches
(6 samples / kit) (July- Dec 2016; Jan- June 2017)
RIQAS Online- Registration, Configuration
Submission and Reports
Submission of Results-Deadline is last Monday of the
Month. Late submission will not be accepted
Result Format and Guidelines new format,
3 indicators: SDI, TS and %DEV
Certificates - COP NEQAS: at least 4 samples per
“cycle” (13B and 14A)
COP Riqas by Randox: at least 6 of 12 samplesFor Queries about the 2016 NEQAS
Program: email us at
lcp_nrl_cc@yahoo.com
call us at (02) 9246101 loc 388
ANNOUNCEMENT:
For NRL updates and notices:
lcp.gov.phFor RIQAS 9128 Service Issues,
please contact:
Marsman Drysdale
Hotline number
(0917) 598-8534You can also read