Measurements of Ion Current from a Corona-needle Charger Using a Faraday Cup Electrometer

 
CONTINUE READING
110                                                                      Chiang Mai J. Sci. 2009; 36(1)

              Chiang Mai J. Sci. 2009; 36(1) : 110-119
              www.science.cmu.ac.th/journal-science/josci.html
              Contributed Paper

Measurements of Ion Current from a Corona-needle
Charger Using a Faraday Cup Electrometer
Panich Intra* [a], and Nakorn Tippayawong [b]
[a] College of Integrated Science and Technology, Rajamangala University of Technology Lanna,
   Chiang Mai 50300, Thailand.
[b] Department of Mechanical Engineering, Faculty of Engineering, Chiang Mai University,
   Chiang Mai 50200, Thailand.
*Author for correspondence; e-mail: panich_intra@yahoo.com, panich.intra@hotmail.com

                                                                            Received: 19 September 2008
                                                                            Accepted: 17 December 2008.

ABSTRACT
          Corona-needle charger is widely used to impose a known net charge distribution on
the aerosol particles for the electrical mobility particle sizer. However, the corona discharge
and charging processes in the corona-needle charger at different operating conditions is not
well understood. In the present paper, measurement of ion current from a corona-needle
charger using a Faraday cup electrometer was performed in order to optimize the corona-
needle charger with respect to maximization of the ion number concentration. It was shown
that the corona onset increased with increasing air flow rate. At higher air flow rate, the ion
current and concentration were found to be relatively high for the same corona voltage. The
highest ion current in the Faraday cup electrometer was found to be about 6.4 × 10-10, and
6.29 × 10-10 A, corresponding to the ion number concentration of about 2.98 × 1013, and 2.93
× 1013 ions/m3 occurring at the corona voltage of 2.9 and 3.7 kV for positive and negative
coronas, and air flow rate at 8.0 L/min, respectively.

Keywords: corona discharge, faraday cup, electrometer, ion current.

1. INTRODUCTION
     One of the most common techniques to              undergo electrical breakdown when the
produce high ion concentrations is corona              electric field strength is high. For the case of
discharge, defined as the low energy electrical        wire and tube, the only place this breakdown
discharge with non-thermal ionization that             can occur is in a very thin layer on the wire
takes place in the vicinity of an electrode of         surface. In this corona region, energy is highly
sufficiently low radius of curvature in a              intense to knock electron from gas molecules
medium the pressure of which is close to               creating positive ions and free electrons.
atmospheric [1]. Corona discharge is produced               There have been numerous studies in
by a non-uniform electrostatic field such as           corona discharge phenomena in the past [2]
that between a needle and plate or a concentric        which are widely used in many industrial
wire and a tube. Air and other gases can               applications such as electrostatic coating and
Chiang Mai J. Sci. 2009; 36(1)                                                                 111

precipitation [3 – 5]. Electrostatic charging of    operating air flow rates and corona voltages.
fine particles by the DC corona dischargers is      The research is aimed at the optimization of
also commonly employed in determining               the corona-needle charger with respect to
particle size distribution by electrical mobility   maximization of the ion concentration. A
technique both corona-wire and corona-              detailed description of the operating principle
needle chargers [6, 7]. Corona-needle charger       of the Faraday cup electrometer was also
is among the most commonly used to                  presented.
generate ions in aerosol diffusion chargers for
the particle sizing instruments. Knowledge of       2. EXPERIMENTAL APPARATUS
the mechanism of the corona discharge               2.1 Corona-needle Charger
attainable in a given charger is needed. The              Figure 1 shows the schematic diagram
reason is that one has to know how particle         of the corona-needle charger used in this study.
charging depends on the ion concentration           The corona-needle charger geometrical
flowing through the charging zone, charging         configuration is similar to the charger used by
time, and the electric field inside it. The issue   Hernandez-Sierra et al. [9], Alonso et al. [10]
of corona discharge in the corona-needle            and Intra and Tippayawong [11]. However,
charger has not been extensively studied in         differences between the present charger and
existing literature. Most papers concern about      existing chargers are aerosol inlet geometry
characteristics of corona discharge in wire to      which was modified to ensure uniform particle
cylinder geometry. Only a few of them focus         distribution across the annular aerosol entrance
on the corona discharge in the needle to nozzle     to charging zone. This charger consists
[8 – 11].                                           essentially of a coaxial needle electrode placed
     In the present paper, the ion current and      along the axis of a cylindrical tube with
number concentration of ions from the               tapered end. The needle electrode is made of
corona-needle charger were measured using           a stainless steel rod, 6 mm in diameter ending
a Faraday cup electrometer at different             in a sharp tip. The angle of the needle cone is

                 Figure 1. Schematic diagram of the corona-needle charger.
112                                                               Chiang Mai J. Sci. 2009; 36(1)

about 10o and the tip radius is about 50 μm,      and a Teflon insulator. To completely shield
as estimated under a microscope. The outer        the HEPA filter collecting the air ions, outer
electrode is made of a stainless steel tube, 30   housing is made of a stainless steel, HEPA
mm in diameter and 15 mm in length with           filter was equipped with a fine collection metal
conical shape. The orifice diameter is about      grid, and was electrically isolated from the
3.5 mm. The distance between the needle           outer housing and ground with Teflon stand
electrode and the cone apex is 1.75 mm. The       (a volume resistivity exceeding 1018 Ωcm). The
corona-needle electrode head is connected to      HEPA filter was used in this work, because
an adjustable DC high voltage supply, while       the collection efficiency for small air ions was
the outer electrode is grounded. The corona       very high. The Faraday cup plays a role to
discharge generates ions which move rapidly       prevent electric noise for measuring low
in the strong corona discharge field towards      electric signal current (in pA range) from
the outer electrode wall.                         accumulated charge of air ions on an internal
                                                  HEPA filter inside the Faraday cup corres-
2.2 Faraday Cup                                   ponding to the total number concentration
     The schematic diagram of the Faraday         of the ions. If the object of measurement is
cup electrometer is shown in Figure 2. It         not shielded completely, noise which is 1000
consists of an outer housing, a High Efficiency   times of resolutions is expected. To transfer
Particulate Air (HEPA) filter, a filter holder,   charges gathered at the HEPA filter to an

                      Figure 2. Schematic diagram of the Faraday cup.
Chiang Mai J. Sci. 2009; 36(1)                                                                  113

electrometer circuit that is outside the Faraday    flowing through a resistor is measured. The
cup, BNC connector is connected to HEPA             circuit adopted two cascaded negative
filter. Because material of HEPA filter is          feedback amplifiers. The extra component in
conductive such as glass fiber, charges collected   this circuit is primarily for fine offset voltage
in the filter can move to the electrometer          adjustment and input/output protection. A
through the low noise cable and BNC                    12V DC power supply capable of
connector without delay. In the case of             providing 100 mA is required. The feedback
existing electrometer air ions flow is curved       capacitor and RC low-pass filter were used
at 90o while air is drifted from sampling probe     to reduce high-frequency noise and to prevent
to the filter. It can become the cause of charge    oscillations of the amplifier output [12]. In
loss. To solve this problem airflow into            order to avoid expensive construction,
Faraday cup is straightened without changing        commercially-available low-cost monolithic
the direction of the flow and loss the charge.      operational amplifiers were used. The
                                                    commercially-available operational amplifiers
2.3 Electrometer Circuit                            used in this circuit is the LMC662, which was
     An electrometer circuit is used to measure     designed for low current measurement
the electric signal current, which are typically    and featured ultra-low input bias current
in the range 1 pA to 1 nA, from the Faraday         (2 fA maximum) and low offset voltage drift
cup. The schematic presentation of an               (1.3 μV/oC) [13]. This circuit gives an output
electrometer circuit design for air ions            voltage of 10 mV per 1 pA of input signal
detection system is shown in Figure 3. This         current. The electrometer circuit was calibrated
circuit is a simple current-to-voltage converter,   with a current injection circuit, high-impedance
where the voltage drop caused by a current          current source [12]. The performance of the

              Figure 3. Schematic diagram of the sensitive electrometer circuit.
114                                                                Chiang Mai J. Sci. 2009; 36(1)

electrometer circuit used in this work was also    Leybold Didactic model 521721, was used
evaluated and compared with a commercial           to maintain the positive and negative corona
electrometer, Keithley model 6517A, and            voltages difference in the charger, generally in
good agreement was found from the                  the range between 1.0 – 5.0 kV. An air sample
comparison [14].                                   was first filtered through a HEPA filter, and
                                                   was then drawn into the charger. The ions
3. EXPERIMENTAL SYSTEM AND PROCEDURE               produced inside the charger are then entered
    The schematic diagram of the experi-           the Faraday cup. In the Faraday cup, the ions
mental system for measurement of ion               were removed from the air stream by the
number concentration from the corona-needle        filter and the resulting ion current flow was
charger is shown in Figure 4. It consists of a     measured with the electrometer. It should be
corona-needle charger, a Faraday cup electro-      noted that the ion current was measured by
meter, a flow system, and a data acquisition       the electrometer corresponding to the ion
and processing system. In our experiments,         number concentration at the charger outlet.
the Faraday cup is connected directly to the       The output signal from the electrometer circuit
charger outlet via a very short connecting pipe.   is in the range of 0 to +10V. It is then sent to
The air flow was regulated and controlled by       the ADAM-4017 analog input module, which
means of a mass flow meter and controller          is a 16-bit, 8 channel analog input module,
with a vacuum pump, typically in the range         controlled and data sampled by an external
between 3.0 – 8.0 L/min. A commercial              personal computer via RS-485 to RS-232
adjustable DC high voltage power supply, a         converter interface. Software running on an

Figure 4. Schematic diagram of the experimental system for measurement of dc ion current
from the corona-needle charger.
Chiang Mai J. Sci. 2009; 36(1)                                                                    115

external computer was developed, based on            needle electrode was measured directly with
Microsoft Visual Basic programming for all           the micro-ampmeter via the outer electrode
data processing. The software is able to display     of the charger. Figure 5 shows the current-
the ion current and number concentration.            voltage characteristics in the charging zone of
The ion current measurements were translated         the charger. In this charger, the corona onset
into ion number concentrations given the total       was found to be about 2.4 kV, and 2.0 kV
air flow rate through the charger. Thus, the         for positive and negative coronas, respectively.
total number concentration of the ion at the         Increase in corona voltage produced a
charger outlet, Ni , can be calculated from the      monotonic increase in charging current. It was
expression [11]                                      shown that the spark-over phenomena
                                                     occurred for both positive and negative
                                              (1)    corona voltages larger than about 4.2 kV.
                                                     Above these values, the current was found to
where Ii is the ion current at the charger outlet,   exhibit a fluctuation in an uncontrollable
e is the elementary charge (1.6 × 10-19 C), and      manner and no measurement could be made.
Qa is the aerosol flow rate.                         Generally, the currents for negative ions were
                                                     slightly higher than those for positive ions. This
4. RESULTS AND DISCUSSION                            was expected because negative ions have
4.1 Current-voltage Characteristics of the           higher electrical mobility than positive ions
Charger                                              ( Z i+ = 1.15 × 10-4 m2/V s, Z i− = 1.425 ×
     The charging current from the corona-           10-4 m2/V s, based on the work of Reischl

Figure 5. Current-voltage characteristics in the charging zone of the corona-needle charger.
116                                                                 Chiang Mai J. Sci. 2009; 36(1)

et al. [15]). Thus, it was more likely to impact    with the same corona voltage. This is because
and deposit on the outer electrode wall of          the ions can be more easily drawn off the
the charger. The ion concentration in the           charger by faster flowing air. In case of
charging zone, N i , of the charger was             positive corona, the ion current and concen-
approximately proportional to the charging          tration appeared to depend on applied
current. Thus, the high ion concentration in        voltage only within a narrow voltage interval.
the charging zone of a charger is desirable         For larger voltages, ion current and
for high particle charging efficiency. The ion      concentration of positive corona became
concentration in the charging zone can be           practically constant, independent of the
estimated from the relation                         applied voltage. Meanwhile, ion current and
                                                    concentration of negative corona slightly
                                             (2)    increases with increasing applied voltage.
                                                    The reason for this may be due to greater
where Zi is the electrical mobility of ions, E      degree of ion loss. It was evident that when
is the electric field, and A is the inner surface   the applied voltage increased, the charging
area of the outer electrode of the charger.         current and electric field strength in the
This charging current increased with the electric   charging zone were found to increase. More
field, hence applied voltage.                       ions have tendency to be electrostatically lost
                                                    in the charging zone of the charger. The ion
4.2 Ion Current and Concentration at the            loss inside the charger due to electrostatic loss
Charger Outlet                                      is defined as the ratio of the ion number
     Figures 6 and 7 show the variations in         concentration at the charger outlet, Nout, over
the ion current and concentration of the            the number concentration of ions inside the
charger outlet with corona voltage at different     charger, Nin. The ion penetration, P, through
operating air flow rates for both positive and      the charger can be estimated by Deutsch-
negative coronas. The resultant ion current and     Anderson equation as [16]
concentration of both positive and negative
coronas were evaluated for 3.0, 5.0, and 8.0                                                     (3)
L/min and 1.0 – 5.0 kV. The obtained results
were expected for the effects of aerosol
flow and corona voltage. As seen in Figure 6,       Form Eq. (3) it can be calculated that the ion
the negative corona onset (i.e. negative ion        penetration was getting smaller with increasing
generation) appeared at about 2.0, 2.1, and         electric field strength as a function of the
2.3 kV for air flow rates of 3.0, 5.0, and 8.0      corona voltage. It is commonly known that
L/min, respectively, while the positive corona      the ion current and concentration for positive
onset was observed at about 2.0, 2.3, and 2.5       corona of the charger was slightly higher than
kV for air flow rates of 3.0, 5.0, 8.0 L/min,       for negative corona. The highest ion current
respectively. For corona voltage less than 2.0      in the Faraday cup was found to be about
kV, the ion current was low. In this range,         6.4 × 10-10, and 6.29 × 10-10 A, corresponding
corona discharge was not present. It can be         to the ion number concentration of about
seen that the corona onset increase with            2.98 × 1013, and 2.93 × 1013 ions/m3 occurring
increasing air flow rate. For both cases, at        at the corona voltage of 2.9, and 3.7 kV for
higher air flow rates, the ion current and          positive and negative coronas, and air flow
concentration were found to be relatively high      rate at 8.0 L/min, respectively.
Chiang Mai J. Sci. 2009; 36(1)                                                          117

                                     (a) Positive corona

                                    (b) Negative corona

        Figure 6. Variation in ion current with corona voltage at the charger outlet.
118                                                          Chiang Mai J. Sci. 2009; 36(1)

                                     (a) Positive corona

                                    (b) Negative corona

Figure 7. Variation in ion number concentration with corona voltage at the charger outlet.
Chiang Mai J. Sci. 2009; 36(1)                                                                    119

5. CONCLUDING REMARKS                                [5] Tan X., and Shang J.K., Electric Field-
     In this paper, the Faraday cup electro-             Induced Intersections of 90o Domain
meter was used to measure the DC ion                     Walls in Tetragonal Ferroelectric Crystals,
current from the corona-needle charger in                Chiang Mai J. Sci., 2005; 32(3): 245 - 252.
order to study the corona discharge inside it.       [6] Intra P., and Tippayawong N., An
                                                         Overview of Aerosol Particle Sensors for
A semi-empirical method based on current
                                                         Size Distribution Measurement, Mj. Int.
measurements was used to determine the total             J. Sci. Tech., 2007; 1(2): 120 - 136.
ion concentration at the outlet of the charger.      [7] Intra P., and Tippayawong N., An
It was found that the corona onset increased             Overview of Differential Mobility
with increasing air flow rate. At higher air flow        Analyzers for Size Classification of
rate, the ion current and concentration were             Nanometer-Sized Aerosol Particles,
found to be relatively high for the same                 Songklanakarin J. Sci. Technol., 2008; 30(2):
corona voltage. The effect of air flow rate              243 - 256.
was more significant than that of corona             [8] Whitby K.T., Generator for Producing
                                                         High Concentration of Small Ions, Rev.
voltage. The negative corona was found to
                                                         Sci. Instrum., 1961; 32(12): 1351-1355.
be in higher concentration than the positive         [9] Hernandez-Sierra A., Alguacil F.J., and
corona. The highest ion current in the Faraday           Alonso M., Unipolar Charging of
cup electrometer was found at the air flow               Nanometer Aerosol Particle in a Corona
rate of 8.0 L/min about 6.4 × 10-10, and 6.29            Ionizer, J. Aerosol Sci., 2003; 34: 733 - 745.
× 10-10 A, for positive and negative coronas,       [10] Alonso M., Martin M.I., and Alguacil
respectively, corresponding to the ion number            F.J., The Measurement of Charging
concentration is about 2.98 × 1013, and 2.93 ×           Efficiencies and Losses of Aerosol
1013 ions/m3 occurring at the corona voltage             Nanoparticles in a Corona Charger, J.
                                                         Electrostatics, 2006; 64: 203 - 214.
of 2.9, and 3.7 kV for positive and negative
                                                    [11] Intra P., and Tippayawong N., Corona
coronas, respectively.                                   Ionizer for Unipolar Diffusion Charging
                                                         of Nanometer Aerosol Particles, 29 th
ACKNOWLEDGEMENT                                          Electrical Engineering Conference, Pattaya,
    Financial support from the Thailand                  Thailand, 2006; 9 - 10 November.
Research Fund (TRF) is gratefully acknow-           [12] Intra P., and Tippayawong N., An Ultra-
ledged.                                                  Low Current Meter for Aerosol
                                                         Detection, CMU. J. Nat. Sci., 2007; 6(2):
REFERENCES                                               313 – 320.
[1] Beuthe T.G., and Chang J.S., Gas Discharge      [13] National Semiconductor Corporation,
    Phenomena, Handbook of Electrostatic                 LMC662 Data Sheet, 2003.
    Processes, Marcel Dekker, New York,             [14] Intra P., and Tippayawong N., An
    1995.                                                Electrostatic Sensor for Nanometer-
[2] Intra P., and Tippayawong N., An                     Sized Aerosol Particles Detection, Asia-
    Over view of Unipolar Charger                        Pacific Symposium on Applied Electromagnetics
    Development for Nano-Aerosols, J.                    and Mechanics, Bangkok, Thailand, 2008;
    Chem. Eng. Japan., in press.                         24 – 25 July.
[3] Parker K.R., Applied Electrostatic              [15] Reischl G.P., Makela J.M., Harch R., and
    Precipitation, Blackie Academic &                    Necid J., Bipolar Charging of Ultrafine
    Professional, New York, 1997.                        Particles in the Size Range below 10 nm,
[4] Lawless P. A., and Sparks L.E., Modeling             J. Aerosol Sci., 1996; 27: 931 - 949.
    Particulate Charging in ESPs, IEEE Trans.       [16] Hinds W.C., Aerosol Technology, John Wiley
    Ind. Applicat., 1988; 24(5): 922-925.                & Sons, New York, 1999.
You can also read